| Alternative title | ORMS-6 : Matériau de Référence Certifié pour le Mercure dans l’Eau |
|---|
| Download | |
|---|
| DOI | Resolve DOI: https://doi.org/10.4224/crm.2025.orms-6 |
|---|
| Author | Search for: Nadeau, Kenny1; Search for: Pihillagawa Gedara, Indumathi1; Search for: LeBlanc, Kelly1ORCID identifier: https://orcid.org/0000-0001-6544-5850; Search for: Brophy, Christine1ORCID identifier: https://orcid.org/0000-0001-9908-9858; Search for: Palmer, Calvin1; Search for: Bordash, Mitchell1; Search for: Yang, Lu1ORCID identifier: https://orcid.org/0000-0002-6896-8603; Search for: Meija, Juris1ORCID identifier: https://orcid.org/0000-0002-3349-5535; Search for: Grinberg, Patricia1ORCID identifier: https://orcid.org/0000-0002-1167-6474 |
|---|
| Affiliation | - National Research Council of Canada
|
|---|
| Format | 3D, Material |
|---|
| Material type | Calibration solution |
|---|
| Subject | Canadian reference material; ORMS-6; mercury; water |
|---|
| Abstract | ORMS-6 is a water spiked with inorganic mercury Certified Reference Material (CRM) from the National Research Council Canada (NRC). The material is packaged in a 50 mL glass ampoule stabilized with 0.5% BrCl. |
|---|
| Analytes | | Analyte | Quantity | Value | Expanded uncertainty | Unit | Type |
|---|
| mercury | mass fraction | 26.18 | 1.10 | pg/g | certified |
|
|---|
| Period of validity | - Valid from 2025-07-08 until 2035-04-30
|
|---|
| Storage conditions | - Temperature: +20 oC
- store at room temperature in dark
|
|---|
| Publication date | 2025-07-01 |
|---|
| Publisher | National Research Council of Canada |
|---|
| Copyright statement | - © 2025
National Research Council of Canada
|
|---|
| Access condition | |
|---|
| References | View items (4)- ‘Evaluation of Measurement Data — Guide to the Expression of Uncertainty in Measurement’, 2008. https://doi.org/10.59161/jcgm100-2008e.
- EPA Method 245.7. Mercury in Water by Cold Vapor Atomic Fluorescence Spectrometry, Revision 2.0 (1 February 2005). https://nepis.epa.gov/Exe/ZyPURL.cgi?Dockey=P1008IY8.txt.
- Veen, Adriaan M. H. van der. ‘Bayesian Analysis of Homogeneity Studies in the Production of Reference Materials’. Accreditation and Quality Assurance 22, no. 6 (2 November 2017): 307–19. https://doi.org/10.1007/s00769-017-1292-6.
- Possolo, Antonio, and Juris Meija. Measurement Uncertainty: A Reintroduction (version 2nd edition). Sistema Interamericano de Metrologia, 2022. https://doi.org/10.4224/1TQZ-B038.
|
|---|
| Export citation | Export as RIS |
|---|
| Collection | Canadian Reference Materials and Methods |
|---|
| Record identifier | 0abf2b4b-dc7f-4253-bb8e-3fafe328a7d2 |
|---|
| Record created | 2025-07-10 |
|---|